» Go to news mainIncidence and severity of paravalvular aortic regurgitation with multidetector computed tomography nominal area oversizing or undersizing after transcatheter heart valve
See full publication
Abstract
OBJECTIVES:
This study sought to compare the influence of the extent of multidetector computed tomography (MDCT) area oversizing on the incidence of paravalvular aortic regurgitation (PAR) between the Sapien 3 and the Sapien XT transcatheter heart valve (THV) to define a new MDCT sizing guideline suitable for the Sapien 3 platform.
BACKGROUND:
The inverse relationship of PAR occurrence and oversizing has been demonstrated for the Sapien XT but the incidence of PAR with comparable oversizing with the Sapien 3 is not known.
METHODS:
Sixty-one prospectively enrolled patients who underwent transcatheter aortic valve replacement with the Sapien 3 THV were compared with 92 patients who underwent transcatheter aortic valve replacement with the Sapien XT THV. Patients were categorized depending on the degree of MDCT area oversizing percentage: undersizing (below 0%), 0% to 5%, 5% to 10%, and above 10%. The primary endpoint was mild or greater PAR on transthoracic echocardiography.
RESULTS:
Mild or greater PAR was present in 19.7% of patients (12 of 61) in the Sapien 3 group and in 54.3% of patients (50 of 92) in the Sapien XT group (p < 0.01). The Sapien 3 group, compared with the Sapien XT group, consistently demonstrated significantly lower rates of mild or greater PAR except for oversizing >10% (p for interaction = 0.54). Moderate or severe PAR rates were also lower in the Sapien 3 group than in the Sapien XT group (3.3% vs. 13.0%, p = 0.04). In the Sapien 3 group, a MDCT area oversizing percentage value of ≤4.17% was identified as the optimal cutoff value to discriminate patients with or without mild or greater PAR.
CONCLUSIONS:
Our retrospective analysis suggests that the Sapien 3 THV displays significantly lower rates of PAR than does the Sapien XT THV. A lesser degree of MDCT area oversizing may be employed for this new balloon-expandable THV.
Copyright © 2015 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Recent News
- ‘A paradigm shift': A first in Canada, donor‑funded nuclear medicine scanner is helping QEII physicians assess and diagnose diseases with precision that wasn’t previously possible — all while tackling waitlists with greater efficiency
- New interventional radiology suite and hybrid operating room at the Halifax Infirmary
- Nova Scotia gets first‑in‑Canada relocatable MRI
- QEII welcomes the first GE StarGuide SPECT‑CT scanner in Canada.
- The first 68Ga‑PSMA PET scans in Atlantic Canada mark a new era in prostate cancer assessment
- Dr. David Volders awarded the Dr. Charles Lo Prize in Radiology Research
- Radiology Research Day 2024
- Program Administrator Award of Excellence