» Go to news mainElectrothermal Coil Detachment Failure in Flow Diverter‑Assisted Coiling of a Small Blister Aneurysm: Technical Considerations and Possible Solutions
Dr. Adela Cora's latest publication.
Abstract
Flow diversion stenting combined with coiling offers both immediate protection from rebleeding for ruptured aneurysms and long-term stability for wide-necked or blister aneurysms. It is particularly useful for tiny ruptured aneurysms, alleviating the concern that small coils may prolapse between the struts of conventional stents.
We employed this technique in a very small, broad-based ruptured aneurysm of the internal carotid, jailing the coiling microcatheter with a Pipeline Embolization Device. However, coil detachment repeatedly failed, until we withdrew the detachment zone into the microcatheter.
We suggest that if the tip of the coiling catheter is adjacent to the stent, contact between the junction zone of the coil and the high metal density of the flow diverter may prevent proper electrothermal coil detachment. Detachment can be undertaken successfully within the microcatheter, though care must be taken thereafter to fully push the detached coil tail into the aneurysm.
Recent News
- QEII welcomes the first GE StarGuide SPECT‑CT scanner in Canada.
- The first 68Ga‑PSMA PET scans in Atlantic Canada mark a new era in prostate cancer assessment
- Dr. David Volders awarded the Dr. Charles Lo Prize in Radiology Research
- Radiology Research Day 2024
- Program Administrator Award of Excellence
- Clinical practice guideline for management of osteoporosis and fracture prevention in Canada: 2023 update
- Dr. Siân Iles interviewed on CBC's Information Morning
- Dr. Steven Beyea and Dr. Adela Cora featured in the latest Beyond Research podcast from ResearchNS
Comments
comments powered by Disqus